How to Calibrate Force Machines According to ASTM E4 ISO 17025 A2LA Accredited

How to Calibrate Force Machines According to ASTM E4 ISO 17025 A2LA Accredited

Frank Bacon Machinery

In this video we calibrate a 100kN Instron 1125 machine. To calibrate force machines we follow the Frank Bacon Procedure that is ISO 17025 accredited under A2LA that follows ASTM E4. In this case we use two different load cells, a 60k standard and 10k standard. We will verify from 250 pounds up to 22,500 pounds. First, we will use the 60k standard to verify points and then use the 10k to get lower points.

Step 1: Apply upper range force and return to no force two times to ensure the cell will return to 0 and there is no memory. 

Step 2: Step through load points as defined by ASTM E4 to verify accuracy throughout the range of the load cell.

Step 3: Ensure the standard and the machine load cells both return to zero after each run.

Step 4: Second run of verification points.

Step 5: Change the standard calibration load cell to verify lower verification points and repeat steps above.

Step 6: Second run of verification points.

Step 7: Record temperature and humidity to identify changes throughput the calibration. 

Step 8: Input verification points into Frank Bacon ISO 17025 workbook to identify any out of tolerance issues and create a calibration report.

Click here to view the step by step process video!

Frank Bacon accredited calibration for tensile and compression, displacement, extensometer and speed

Click here for our A2LA calibration scope accreditation

Frank Bacon Machinery Sales Co. has The American Association for Laboratory Accreditation (A2LA) accreditation to ISO/IEC 17025 for calibration services. Our American Association for Laboratory Accreditation (A2LA) is a proud achievement and provides our clients confidence in our calibration services.

Adherence to these standards provides our customers with confidence that each machine is receiving excellent calibration service. Frank Bacon understands the importance of precision accuracy when operating your equipment. The quality of calibration Frank Bacon provides ensures your equipment may run optimally.

A2LA Mechanical calibration services offered by Frank Bacon:

Dimensional:

Linear Displacement – Encoder

Linear Displacement – Digital Indicator

Extensometer

Mechanical:

Force – Force Machines – Tension and Compression

Time & Frequency:

Speed – Linear Travel

We adhere to ASTMs including but not limited to:

               ASTM E2309, ASTM E83, ASTM E4, ASTM E2658

What sets Frank Bacon apart from competitors and calibration houses?

Frank Bacon has numerous highly trained technicians that can travel across the country to perform calibrations to the highest standards. Unique to Frank Bacon, along with being an accredited calibration house, we are also expert machine builders that have been upgrading and retrofitting machines since 1953.  Because of our experience with machine building, we can do much more than a general calibration company, we can repair, troubleshoot, and upgrade your equipment as needed.

Click here to receive a calibration quote!

Contact Frank Bacon Today!

If you have any questions about our calibration services please contact a sales engineer at Frank Bacon. Our experts will provide you with the information and services right for you!

Phone: 586.756.4280

Toll Free: 800- 394-4280

sales@frankbacon.com

Click here to view our Tensile Tester Inventory

Click here to view our YouTube Channel

Added to Basket

Proceed To Quote